Sökning: id:"swepub:oai:DiVA.org:liu-147514" >
5-ALA fluorescence ...
5-ALA fluorescence and laser Doppler flowmetry for guidance in a stereotactic brain tumor biopsy
-
- Haj-Hosseini, Neda, 1980- (författare)
- Linköpings universitet,Avdelningen för medicinsk teknik,Tekniska fakulteten,MINT
-
- Richter, Johan, 1960- (författare)
- Linköpings universitet,Institutionen för medicinsk teknik,Tekniska fakulteten,Region Östergötland, Neurokirurgiska kliniken US
-
- Milos, Peter, 1969- (författare)
- Linköpings universitet,Avdelningen för neuro- och inflammationsvetenskap,Medicinska fakulteten,Region Östergötland, Neurokirurgiska kliniken US,MINT
-
visa fler...
-
- Hallbeck, Martin, 1970- (författare)
- Linköpings universitet,Avdelning för neurobiologi,Medicinska fakulteten,Region Östergötland, Klinisk patologi
-
- Wårdell, Karin, 1959- (författare)
- Linköpings universitet,Avdelningen för medicinsk teknik,Tekniska fakulteten,MINT
-
visa färre...
-
(creator_code:org_t)
- Optical Society of America, 2018
- 2018
- Engelska.
-
Ingår i: Biomedical Optics Express. - : Optical Society of America. - 2156-7085. ; 9:5, s. 2284-2296
- Relaterad länk:
-
https://liu.diva-por... (primary) (Raw object)
-
visa fler...
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- A fiber optic probe was developed for guidance during stereotactic brain biopsy procedures to target tumor tissue and reduce the risk of hemorrhage. The probe was connected to a setup for the measurement of 5-aminolevulinic acid (5-ALA) induced fluorescence and microvascular blood flow. Along three stereotactic trajectories, fluorescence (n = 109) and laser Doppler flowmetry (LDF) (n = 144) measurements were done in millimeter increments. The recorded signals were compared to histopathology and radiology images. The median ratio of protoporphyrin IX (PpIX) fluorescence and autofluorescence (AF) in the tumor was considerably higher than the marginal zone (17.3 vs 0.9). The blood flow showed two high spots (3%) in total. The proposed setup allows simultaneous and real-time detection of tumor tissue and microvascular blood flow for tracking the vessels.
Ämnesord
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering (hsv//eng)
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas