Mechanical chest compressions improved aspects of CPR in the LINC trial

Esibov, Alexander (author)
Banville, Isabelle (author)
Chapman, Fred W. (author)
show more...
Boomars, Rene (author)
Box, Martyn (author)
Rubertsson, Sten (author)
Uppsala universitet,Anestesiologi och intensivvård
show less...
 (creator_code:org_t)
Elsevier BV, 2015
2015
English.
In: Resuscitation. - : Elsevier BV. - 0300-9572 .- 1873-1570. ; 91, s. 116-121
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Aim: We studied resuscitation process metrics in patients with out-of-hospital cardiac arrest enrolled in a randomized trial comparing one protocol designed to best use a mechanical CPR device, with another based on the 2005 European Resuscitation Council guidelines for manual CPR. Methods: We analyzed clinical data, ECG signals, and transthoracic impedance signals for a subset of the patients in the LUCAS in Cardiac Arrest (LINC) trial, including 124 patients randomized to mechanical and 82 to manual CPR. Chest compression fraction (CCF) was defined as the fraction of time during cardiac arrest that chest compressions were administered. Results: Patients in the mechanical CPR group had a higher CCF than those in the manual CPR group [0.84 (0.78, 0.91) vs. 0.79 (0.70, 0.86), p < 0.001]. The median duration of their pauses for defibrillation was also shorter [0 s (0, 6.0) vs. 10.0 s (7.0, 14.3), p < 0.001]. Compressions were interrupted for a median of 36.0 s to apply the compression device. There was no difference between groups in duration of the longest pause in compressions [32.5 s vs. 26.0 s, p = 0.24], number of compressions received per minute [86.5 vs. 88.3, p = 0.47], defibrillation success rate [73.2% vs. 81.0%, p = 0.15], or refibrillation rate [74% vs. 77%, p = 0.79]. Conclusions: A protocol using mechanical chest compression devices reduced interruptions in chest compressions, and enabled defibrillation during ongoing compressions, without adversely affecting other resuscitation process metrics. Future emphasis on optimizing device deployment may be beneficial.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Anestesi och intensivvård (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Anesthesiology and Intensive Care (hsv//eng)

Keyword

Cardiopulmonary resuscitation (CPR)
Mechanical CPR
Out-of-hospital cardiac arrest (OHCA)
Chest compression fraction
Perishock pause
Defibrillation

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view