SwePub
Sök i LIBRIS databas

  Utökad sökning

id:"swepub:oai:gup.ub.gu.se/305125"
 

Sökning: id:"swepub:oai:gup.ub.gu.se/305125" > A model for evaluat...

A model for evaluating the use of imaging in image-guided interventional procedures-possible implications on optimisation of radiation protection.

Lundh, Charlotta, 1977 (författare)
University of Gothenburg,Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för medicinsk strålningsvetenskap,Institute of Clinical Sciences, Department of Medical Radiation Sciences,Sahlgrenska University Hospital
Ivarsson, Jonas, 1976 (författare)
University of Gothenburg,Gothenburg University,Göteborgs universitet,Institutionen för tillämpad informationsteknologi (GU),Department of Applied Information Technology (GU)
Falkenberg, Mårten, 1959 (författare)
University of Gothenburg,Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
visa fler...
Båth, Magnus, 1974 (författare)
University of Gothenburg,Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för medicinsk strålningsvetenskap,Institute of Clinical Sciences, Department of Medical Radiation Sciences,Sahlgrenska University Hospital
Almén, A (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,LUCC: Lunds universitets cancercentrum,Övriga starka forskningsmiljöer,Medical Radiation Physics, Malmö,Lund University Research Groups,LUCC: Lund University Cancer Centre,Other Strong Research Environments,Swedish Radiation Safety Authority
visa färre...
 (creator_code:org_t)
2021-04-19
2021
Engelska.
Ingår i: Radiation protection dosimetry. - : Oxford University Press (OUP). - 1742-3406 .- 0144-8420. ; 195:3-4, s. 139-144
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The present study focuses on introducing the concept of optimisation and proposing a model, including evaluation of image quality, to be used in the clinical routines where image-guided intervention is being performed. The overall aim of the study was to develop a model for evaluating the use of imaging in X-ray-guided interventional procedures and its possible implications on optimisation of radiation protection. In the search for an adequate evaluation model, data from endovascular interventions of the aorta (EVAR procedures) were used. The procedure was schematically described in steps. Every imaging event was connected to the steps in the medical procedure and was also described with the purpose of the imaging event. Available technical, as well as procedural parameters, were studied and analysed. Data were collected from the X-ray equipment for 70 EVAR procedures and, out of these, 12 procedures were randomly selected to be recorded on video to understand the procedure better. It was possible to describe the EVAR procedures in a general way with explanations of the clinical purpose connected to each imaging event. Possible quality parameters of the procedure were identified for the imaging events (radiation dose, image quality). The model method still needs to be refined and will then be applied to clinical data and to other clinical procedures to test the validity.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.