Limitations of a pencil beam approach to photon dose calculations in the head and neck region

Hurkmans, Coen (author)
Knöös, Tommy (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Nilsson, Per (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
show more...
Svahn-Tapper, Gudrun (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Danielsson, Hans (author)
show less...
 (creator_code:org_t)
1995
1995
English.
In: Radiotherapy and Oncology. - 1879-0887. ; 37:1, s. 74-80
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • The inherent limitations of a specific pencil beam model have been studied when applied to a cylindrical geometry simulating the neck region. A comparison is made between measured and calculated absorbed dose in a cylindrical phantom. The goal is to quantify the deviations in the absorbed dose level, i.e., the dose per monitor unit, when photons are used for the treatment of head and neck tumours. Square fields ranging from 5 x 5 up to 30 x 30 cm2 are studied for photon beam energies of 60Co, 4, 6 and 18 MV. Ionisation chamber measurements have been performed in the cylinder as well as in two other configurations in order to trace the origin of possible deviations. For 18 MV no significant deviations are found between measurement and calculation in the cylindrical configuration. For the lower energies, an overestimation of the calculated dose in the cylindrical configuration up to about 6% for a 20 x 20-cm2 60Co field has been found. These deviations have been traced to the basic approximation for the integration volume for phantom scatter calculations inherent in this pencil beam implementation.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Keyword

Radiotherapy
Dose planning
Pencil beam algorithm
Head and neck region
Patient modelling

Publication and Content Type

art (subject category)
ref (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Find more in SwePub

By the author/editor
Hurkmans, Coen
Knöös, Tommy
Nilsson, Per
Svahn-Tapper, Gu ...
Danielsson, Hans
About the subject
MEDICAL AND HEALTH SCIENCES
MEDICAL AND HEAL ...
and Clinical Medicin ...
and Radiology Nuclea ...
Articles in the publication
Radiotherapy and ...
By the university
Lund University

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view