SwePub
Sök i LIBRIS databas

  Utökad sökning

id:"swepub:oai:lup.lub.lu.se:b8634e4d-befa-42aa-b062-2e3ce6d04138"
 

Sökning: id:"swepub:oai:lup.lub.lu.se:b8634e4d-befa-42aa-b062-2e3ce6d04138" > Biosensor based on ...

Biosensor based on cellobiose dehydrogenase for detection of catecholamines

Stoica, Leonard (författare)
Lund University,Lunds universitet,Centrum för analys och syntes,Kemiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Centre for Analysis and Synthesis,Department of Chemistry,Departments at LTH,Faculty of Engineering, LTH
Lindgren-Sjolander, A (författare)
Ruzgas, Tautgirdas (författare)
Lund University,Lunds universitet,Centrum för analys och syntes,Kemiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Centre for Analysis and Synthesis,Department of Chemistry,Departments at LTH,Faculty of Engineering, LTH
visa fler...
Gorton, Lo (författare)
Lund University,Lunds universitet,Centrum för analys och syntes,Kemiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Centre for Analysis and Synthesis,Department of Chemistry,Departments at LTH,Faculty of Engineering, LTH
visa färre...
 (creator_code:org_t)
2004-06-29
2004
Engelska.
Ingår i: Analytical Chemistry. - : American Chemical Society (ACS). - 1520-6882 .- 0003-2700. ; 76:16, s. 4690-4696
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • A cellobiose dehydrogenase (CDH)-modified graphite electrode was designed for amperometric detection of catecholamines in the flow injection mode, by their recycling between the graphite electrode (+300 mV vs AgAgCl) and the reduced FAD cofactor of adsorbed CDH, resulting in an amplified response signal. The high efficiency of the enzyme-catecholamine reaction leads to a detection limit below 1 nM and a sensitivity of 15.8 A(.)M(-1.)cm(2) 2 (1150 nA/muM) for noradrenaline, with a coverage of less than 2.5 mug of CDH adsorbed on the electrode surface (0.073 cm(2)). Working parameters such as pH, cellobiose concentration, carrier buffer, and applied potential were optimized, using hydroquinone as a model analyte. The sensitivity, linear range, and amplification factor can be modulated by the steady-state concentration of cellobiose in the flow buffer. The response of the sensor decreases only 2% when run continuously for 4 h in the flow injection mode. The response peak maximum is obtained within 6 s at a flow rate of 0.5 mL/min, representing the time of the entire sample segment to pass the electrode. CDH enzymes from Phanerochaete chrysosporium and Sclerotium rolfsii were investigated, providing different characteristics of the sensor, with sensors made with CDH from P. chrysosporium being the better ones.

Ämnesord

NATURVETENSKAP  -- Kemi -- Analytisk kemi (hsv//swe)
NATURAL SCIENCES  -- Chemical Sciences -- Analytical Chemistry (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Stoica, Leonard
Lindgren-Sjoland ...
Ruzgas, Tautgird ...
Gorton, Lo
Om ämnet
NATURVETENSKAP
NATURVETENSKAP
och Kemi
och Analytisk kemi
Artiklar i publikationen
Analytical Chemi ...
Av lärosätet
Lunds universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy