Evaluation of an algorithm for arterial lumen diameter measurements by means of ultrasound.

Cinthio, Magnus (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Jansson, Tomas (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Eriksson, Anders (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
visa fler...
Rydén Ahlgren, Åsa (författare)
Lund University,Lunds universitet,Klinisk fysiologi och nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,Clinical Physiology and Nuclear Medicine, Malmö,Lund University Research Groups
Persson, Hans W (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Lindström, Kjell (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
visa färre...
 (creator_code:org_t)
2010-07-16
2010
Engelska.
Ingår i: Medical & Biological Engineering & Computing. - : Springer Science and Business Media LLC. - 0140-0118 .- 1741-0444. ; 48, s. 1133-1140
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • We have developed an algorithm for arterial luminal diameter measurement by means of ultrasound and evaluated the algorithm on agar vessel phantoms and in vivo. The algorithm utilises relative threshold detection on the inner slopes of the arterial walls before the resolution is improved by solving the equation of a straight line between the samples around the threshold value. Further, correction distances added to compensate for the underestimation when using the inner slopes were found to be 304 mum for the near wall and 415 mum for the far wall. The measured mean diameters of ten consecutive images of 3-, 6- and 9-mm phantoms were 3,006 mum (SD 4), 5,918 mum (SD 1) and 8,936 mum (SD 2), respectively. The mean differences between the images were 0.19, 0.04 and 0.37 mum, respectively. In vivo, the intra- and inter-observer variabilities were -64 mum (2SD 358) and -57 mum (2SD 366), respectively.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Annan medicinteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Other Medical Engineering (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy