Reliable phase-contrast flow volume magnetic resonance measurements are feasible without adjustment of the velocity encoding parameter

Lagerstrand, Kerstin M (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Svensson, Frida, 1979 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Polte, Christian Lars (författare)
Gothenburg University,Göteborgs universitet,Institutionen för medicin, avdelningen för molekylär och klinisk medicin,Institute of Medicine, Department of Molecular and Clinical Medicine
visa fler...
Bech-Hanssen, Odd, 1956 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för medicin, avdelningen för molekylär och klinisk medicin,Institute of Medicine, Department of Molecular and Clinical Medicine
Starck, Göran (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Chodorowski, Artur, 1959 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
Johnsson, Åse (Allansdotter), 1966 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
visa färre...
 (creator_code:org_t)
2020
2020
Engelska.
Ingår i: Journal of Medical Imaging. - 2329-4310 .- 2329-4302. ; 7:6
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Purpose: To show that adjustment of velocity encoding (VENC) for phase-contrast (PC) flow volume measurements is not necessary in modern MR scanners with effective background velocity offset corrections. Approach: The independence on VENC was demonstrated theoretically, but also experimentally on dedicated phantoms and on patients with chronic aortic regurgitation (n = 17) and one healthy volunteer. All PC measurements were performed using a modern MR scanner, where the pre-emphasis circuit but also a subsequent post-processing filter were used for effective correction of background velocity offset errors. Results: The VENC level strongly affected the velocity noise level in the PC images and, hence, the estimated peak flow velocity. However, neither the regurgitant blood flow volume nor the mean flow velocity displayed any clinically relevant dependency on the VENC level. Also, the background velocity offset was shown to be close to zero (

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk laboratorie- och mätteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Laboratory and Measurements Technologies (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Maskinteknik -- Strömningsmekanik och akustik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Mechanical Engineering -- Fluid Mechanics and Acoustics (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Image Processing (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

velocity encoding
phase contrast
background velocity offset
velocity-to-noise-ratio
magnetic resonance imaging

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy