Assessment of horizontal laminar air flow instrument table for additional ultraclean space during surgery

Nilsson, Kjell G (author)
Umeå universitet,Ortopedi
Lundholm, Rolf (author)
Umeå universitet,Institutionen för klinisk mikrobiologi
Friberg, Sven (author)
Umeå universitet,Ortopedi
 (creator_code:org_t)
Elsevier, 2010
2010
English.
In: Journal of Hospital Infection. - : Elsevier. - 0195-6701 .- 1532-2939. ; 76:3, s. 243-246
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • The area in a vertical ultraclean laminar air flow (LAF) theatre is usually too small to accommodate all the equipment needed for major surgery. We investigated the addition of an instrument table supplied with fixed ultraclean LAF and placed alongside the existing main LAF unit, to determine its physical and bacteriological effect on the main unit. In phase 1, with two investigators but without a patient, smoke tests showed no intrusion of air from the table into the main unit and particle counts did not show any adverse effect on the main LAF unit. In phase 2, during patients undergoing two total knee replacements, the LAF table and a table without LAF were placed alongside the main LAF unit. The tables were subjected to the activity of an extra operating room (OR) nurse working from inside the main LAF vigorously simulating handling of instruments. During this activity, the >5 μm particle counts were 275/m3 at the instrument table with LAF and 8550/m3 at the table without LAF (P < 0.0001). Also, without the OR nurse activity, the particle counts, just inside the main unit and adjacent to the LAF table, were significantly reduced (P < 0.03–0.003). Sedimentation plates on the LAF table and in the main unit registered 22 and 25 cfu/m2/h respectively compared with 45 cfu/m2/h at the instrument table without LAF. In conclusion, the results from the smoke tests, particle counts and bacteriological evaluation showed that the additional instrument table supplied with LAF is efficient and can be safely used as an extension additional to a main OR LAF unit.

Keyword

Air contamination
Colony count
Laminar air flow
Operating room
Particle count

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Find more in SwePub

By the author/editor
Nilsson, Kjell G
Lundholm, Rolf
Friberg, Sven
Articles in the publication
Journal of Hospi ...
By the university
Umeå University

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view