SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Magnusson O)
 

Sökning: WFRF:(Magnusson O) > (1990-1999) > Improvements in abs...

Improvements in absorbed dose measurements for external radiation therapy using ferrous dosimeter gel and MR imaging (FeMRI)

Bäck, Sven (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Magnusson, Peter (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Fransson, A (författare)
visa fler...
Olsson, L E (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Montelius, A (författare)
Uppsala universitet,Institutionen för onkologi, radiologi och klinisk immunologi,Sjukhusfysik
Holmberg, O (författare)
St Luke's Hospital
Andreo, P (författare)
Karolinska Institutet
Mattsson, S (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
visa färre...
 (creator_code:org_t)
1999-01-01
1998
Engelska 16 s.
Ingår i: Physics in Medicine and Biology. - : IOP Publishing. - 0031-9155 .- 1361-6560. ; 43, s. 261-
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • A ferrous gel, based on ferrous (Fe) sulphate and agarose, was used with a clinical magnetic resonance imaging (MRI) scanner to obtain relative dose distribution data from therapeutic photon and electron beams. The FeMRI gel was scanned using a new MRI acquisition protocol optimized for T1 measurements. Thorough comparisons with silicon semiconductor detector and ionization chamber measurements, as well as with Monte Carlo calculations, were performed in order to quantify the improvements obtained using FeMRI for dose estimations. Most of the relative doses measured with FeMRI were within 2% of the doses measured with other methods. The larger discrepancies (2-4%) found at shallow depths are discussed. The uncertainty in relative dose measurements using FeMRI was significantly improved compared with previously reported results (5-10%, one standard deviation, 1 SD), and is today between 1.6% and 3.3% (depending on dose level, 2 SD). This corresponds to an improvement in the minimum detectable dose (3 SD above background) from approximately 2 Gy to better than 0.6 Gy. The results obtained in this study emphasize the importance of obtaining basic FeMRI dose data before the method is extended to complicated treatment regimes.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Electrons
Ferrous Compounds
Gels
Magnetic Resonance Imaging
Phantoms, Imaging
Photons
Radiotherapy Dosage
Radiotherapy Planning, Computer-Assisted
Radiotherapy, High-Energy
Reproducibility of Results
Sepharose

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy