Novel, clinically applicable method to measure step-width during the swing phase of gait
-
- Bäcklund, Tomas (författare)
- Umeå universitet,Radiofysik
-
- Öhberg, Fredrik, 1969- (författare)
- Umeå universitet,Radiofysik
-
- Johansson, Gudrun (författare)
- Umeå universitet,Institutionen för samhällsmedicin och rehabilitering
-
visa fler...
-
- Grip, Helena (författare)
- Umeå universitet,Radiofysik
-
- Sundström, Nina (författare)
- Umeå universitet,Radiofysik
-
visa färre...
-
(creator_code:org_t)
- 2020-06-30
- 2020
- Engelska.
-
Ingår i: Physiological Measurement. - : Institute of Physics Publishing (IOPP). - 0967-3334 .- 1361-6579. ; 41:6
- Relaterad länk:
-
https://doi.org/10.1...
-
visa fler...
-
https://umu.diva-por... (primary) (Raw object)
-
https://doi.org/10.1...
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Objective: Step-width during walking is an indicator of stability and balance in patients with neurological disorders, and development of objective tools to measure this clinically would be a great advantage. The aim of this study was to validate an in-house-developed gait analysis system (Striton), based on optical and inertial sensors and a novel method for stride detection, for measuring step-width during the swing phase of gait and temporal parameters.Approach: The step-width and stride-time measurements were validated in an experimental setup, against a 3D motion capture system and on an instrumented walkway. Further, test-retest and day-to-day variability were evaluated, and gait parameters were collected from 87 elderly persons (EP) and four individuals with idiopathic normal pressure hydrocephalus (iNPH) before/after surgery.Main results: Accuracy of the step-width measurement was high: in the experimental setup mean error was 0.08 +/- 0.25 cm (R = 1.00) and against the 3D motion capture system 0.04 +/- 1.12 cm (R = 0.98). Test-retest and day-to-day measurements were equal within +/- 0.5 cm. Mean difference in stride time was -0.003 +/- 0.008 s between Striton and the instrumented walkway. The Striton system was successfully applied in the clinical setting on individuals with iNPH, which had larger step-width (6.88 cm, n = 4) compared to EP (5.22 cm, n = 87).Significance: We conclude that Striton is a valid, reliable and wearable system for quantitative assessment of step-width and temporal parameters during gait. Initial measurements indicate that the newly defined step-width parameter differs between EP and patients with iNPH and before/after surgery. Thus, there is potential for clinical applicability in patients with reduced gait stability.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Medicinska och farmaceutiska grundvetenskaper -- Neurovetenskaper (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Basic Medicine -- Neurosciences (hsv//eng)
Nyckelord
- gait
- swing phase
- step-width
- inertial measurement unit
- wearable
- optical sensor