Determining Heart Activity Present in the Pressure Sensors of a Dialysis Machine

Holmer, Mattias (author)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Grigonyte, Egle (author)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Solem, Kristian (author)
Lund University,Lunds universitet,Institutioner vid LTH,Lunds Tekniska Högskola,Departments at LTH,Faculty of Engineering, LTH
show more...
Olde, Bo (author)
Sandberg, Frida (author)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Sörnmo, Leif (author)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Murray, Alan (editor)
show less...
 (creator_code:org_t)
2013
2013
English.
In: Computing in Cardiology. - 2325-8861. ; , s. 217-220
  • Conference paper (peer-reviewed)
Abstract Subject headings
Close  
  • Determination of heart status during dialysis can im- prove patient monitoring. Pressure sensors in the dialysis machine measures the heart pulses that propagates in the body and enter the extracorporeal blood circuit. A peri- staltic blood pump, located in the same circuit, introduces strong periodic pressure pulses that interfere with the much weaker cardiac component. These signal characteristics make the extraction of the heart activity challenging. In the present study, we explore the possibility to extract and analyze the cardiac component using simulated data. The accuracy of the timing of each heartbeat is analyzed. Ad- ditionally, the heart component is extracted from patient pressure recordings, and compared to the heart rate com- puted from a photoplethysmogram. The results show that heart timings can be accurately determined using the pres- sure sensors of a dialysis machine.

Subject headings

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering (hsv//eng)

Publication and Content Type

kon (subject category)