- Surface Electromyography (sEMG) has applications in prosthetics, diagnostics and neuromuscular rehabilitation. Self-adhesive Ag/AgCl are the electrodes preferentially used to capture sEMG in short-term studies, however their long-term application is limited. In this study we designed and evaluated a fully integrated smart textile band with electrical connecting tracks knitted with intarsia techniques and knitted textile electrodes. Real-time myoelectric pattern recognition for motor volition and signal-to-noise ratio (SNR) were used to compare its sensing performance versus the conventional Ag-AgCl electrodes. After a comprehending measurement and performance comparison of the sEMG recordings, no significant differences were found between the textile and the Ag-AgCl electrodes in SNR and prediction accuracy obtained from pattern recognition classifiers.
Ämnesord
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Annan medicinteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Other Medical Engineering (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk laboratorie- och mätteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Laboratory and Measurements Technologies (hsv//eng)
Nyckelord
- Electrodes
- Real-time systems
- Pattern recognition
- Signal to noise ratio
- Electromyography
- Fabrics
- Applied Medical Technology
- Tillämpad medicinsk teknik
Publikations- och innehållstyp
- ref (ämneskategori)
- kon (ämneskategori)