A flow injection voltammetric electronic tongue applied to paper mill industrial waters

Gutes, A. (författare)
Gutés, A., Sensors and Biosensors Group, Department of Chemistry, Autonomous University of Barcelona, Edifici Cn, 08193 Bellaterra, Spain
Cespedes, F. (författare)
Sensors and Biosensors Group, Department of Chemistry, Autonomous University of Barcelona, Edifici Cn, 08193 Bellaterra, Spain
del, Valle M. (författare)
del Valle, M., Sensors and Biosensors Group, Department of Chemistry, Autonomous University of Barcelona, Edifici Cn, 08193 Bellaterra, Spain
visa fler...
Louthander, D. (författare)
Krantz-Rülcher, Christina (författare)
Linköpings universitet,Tekniska högskolan,Tillämpad Fysik
Winquist, Fredrik (författare)
Linköpings universitet,Tekniska högskolan,Tillämpad Fysik
visa färre...
Gutés, A, Sensors and Biosensors Group, Department of Chemistry, Autonomous University of Barcelona, Edifici Cn, 08193 Bellaterra, Spain Sensors and Biosensors Group, Department of Chemistry, Autonomous University of Barcelona, Edifici Cn, 08193 Bellaterra, Spain (creator_code:org_t)
Elsevier BV, 2006
2006
Engelska.
Ingår i: Sensors and actuators. B, Chemical. - : Elsevier BV. - 0925-4005 .- 1873-3077. ; 115:1, s. 390-395
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • A voltammetric electronic tongue with automated operation based on the flow injection (FIA) technique was applied to the characterization of wastewaters coming from the paper mill industry. A metallic multielectrode array - formed by platinum, gold and rhodium electrodes - was employed as the detection system, while the measurements were based on large amplitude pulse voltammetry (LAPV). LAPV consisted in scans of pulses from to 0 to 1.8 V at 0.2 V steps. Five current values were recorded for each pulse, so a set of 300 current values (three electrodes × 20 pulses × five values) was recorded for each sample. Samples were first discriminated using Principal Component Analysis (PCA), while Artificial Neural Networks were used for the characterization and prediction of chemical oxygen demand, conductivity and pH. The system may be used for the quick identification and monitoring of the quality of used waters in these industrial facilities. © 2005 Elsevier B.V. All rights reserved.

Nyckelord

Chemical oxygen demand
Electronic tongue
Flow injection analysis
Paper mill industry
Voltammetry
TECHNOLOGY
TEKNIKVETENSKAP

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy