SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Lundström T)
 

Sökning: WFRF:(Lundström T) > (1995-1999) > Surface plasmon res...

Surface plasmon resonance (SPR) analysis of coagulation in whole blood with application in prothrombin time assay

Hansson, Kenny, 1972- (författare)
Linköpings universitet,Klinisk kemi,Hälsouniversitetet
Vikinge, T. P. (författare)
Linköpings universitet,Tillämpad Fysik,Hälsouniversitetet
Rånby, M. (författare)
Linköpings universitet,Klinisk kemi,Hälsouniversitetet
visa fler...
Tengvall, Pentti, 1951- (författare)
Linköpings universitet,Tillämpad Fysik,Tekniska högskolan
Lundström, Ingemar, 1941- (författare)
Linköpings universitet,Tillämpad Fysik,Tekniska högskolan
Johansen, Knut (författare)
Linköpings universitet,Tillämpad Fysik,Hälsouniversitetet
Lindahl, Tomas, 1954- (författare)
Linköpings universitet,Klinisk kemi,Hälsouniversitetet
visa färre...
 (creator_code:org_t)
1999
1999
Engelska.
Ingår i: Biosensors & bioelectronics. - 0956-5663 .- 1873-4235. ; 14:8-9, s. 671-682
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • It is previously shown that surface plasmon resonance (SPR) can be used to study blood plasma coagulation. This work explores the use of this technique for the analysis of tissue factor induced coagulation, i.e. prothrombin time (PT) analysis, of whole blood and plasma. The reference method was nephelometry. The prothrombin time analysis by SPR was performed by mixing two volumes of blood/plasma, one volume of thromboplastin, and one volume of CaCl2 solution directly on a sensor surface. The measurements show good agreement between nephelometry and SPR plasma analysis and also between SPR plasma and whole blood analysis. The effect of anticoagulant treatment on the clotting times was significant both quantitatively and qualitatively. The impact on the SPR signal of different physiological events in the coagulation process is discussed, and tentative interpretations of the sensorgram features are given. The major advantage of the SPR method compared to nephelometry is the possibility to perform analysis on whole blood instead of plasma. In conclusion, SPR is a promising method for whole blood coagulation analysis.

Nyckelord

MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy