SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Schmidt Jan)
 

Sökning: WFRF:(Schmidt Jan) > (1990-1999) > Serum cystatin C, d...

Serum cystatin C, determined by a rapid, automated particle-enhanced turbidimetric method, is a better marker than serum creatinine for glomerular filtration rate

Kyhse-Andersen, Jan (författare)
Schmidt, Camilla (författare)
Nordin, Gunnar (författare)
visa fler...
Andersson, Britta (författare)
Nilsson-Ehle, Peter (författare)
Lund University,Lunds universitet,Avdelningen för klinisk kemi och farmakologi,Institutionen för laboratoriemedicin,Medicinska fakulteten,Division of Clinical Chemistry and Pharmacology,Department of Laboratory Medicine,Faculty of Medicine
Lindström, Veronica (författare)
Lund University,Lunds universitet,Avdelningen för klinisk kemi och farmakologi,Institutionen för laboratoriemedicin,Medicinska fakulteten,Division of Clinical Chemistry and Pharmacology,Department of Laboratory Medicine,Faculty of Medicine
Grubb, Anders (författare)
Lund University,Lunds universitet,Avdelningen för klinisk kemi och farmakologi,Institutionen för laboratoriemedicin,Medicinska fakulteten,Division of Clinical Chemistry and Pharmacology,Department of Laboratory Medicine,Faculty of Medicine
visa färre...
 (creator_code:org_t)
1994
1994
Engelska.
Ingår i: Clinical Chemistry. - 0009-9147. ; 40:10, s. 1921-1926
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • We describe a fully automated particle-enhanced turbidimetric assay for cystatin C in undiluted serum and EDTA-plasma. The throughput is 90 samples per hour and urgent samples can be analyzed in 7 min. The assay range (0.4-14.1 mg/L) covers the concentration range in health and disease. The within- and between-run imprecision is 0.9% and 2.2%, respectively. Analytical recovery of additions of recombinant cystatin C averaged 98%. Rheumatoid factors (< or = 323,000 IU/L), bilirubin (< or = 150 mumol/L), hemoglobin (< or = 1.2 g/L), and triglycerides (< or = 8.5 mmol/L) do not interfere in the assay. In view of the superior (by ROC analysis) diagnostic accuracy of serum concentrations of cystatin C for reduced glomerular filtration rate (GFR) in comparison with creatinine, cystatin C seems an attractive alternative to creatinine for estimation of GFR.

Ämnesord

NATURVETENSKAP  -- Biologi -- Biokemi och molekylärbiologi (hsv//swe)
NATURAL SCIENCES  -- Biological Sciences -- Biochemistry and Molecular Biology (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy