SwePub
Sök i LIBRIS databas

  Utökad sökning

onr:"swepub:oai:DiVA.org:uu-261991"
 

Sökning: onr:"swepub:oai:DiVA.org:uu-261991" > In vitro biomechani...

In vitro biomechanical evaluation of four fixation techniques for distractive-flexion injury stage 3 of the cervical spine

Henriques, Thomas (författare)
Cunningham, Bryan W. (författare)
Mcafee, Paul C. (författare)
visa fler...
Olerud, Claes (författare)
Uppsala universitet,Ortopedi
visa färre...
 (creator_code:org_t)
2015-03-06
2015
Engelska.
Ingår i: Upsala Journal of Medical Sciences. - : Uppsala Medical Society. - 0300-9734 .- 2000-1967. ; 120:3, s. 198-206
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Purpose. Anterior plate fixation has been reported to provide satisfactory results in cervical spine distractive flexion (DF) injuries stages 1 and 2, but will result in a substantial failure rate in more unstable stage 3 and above. The aim of this investigation was to determine the biomechanical properties of different fixation techniques in a DF-3 injury model where all structures responsible for the posterior tension band mechanism are torn. Methods. The multidirectional three-dimensional stiffness of the subaxial cervical spine was measured in eight cadaveric specimens with a simulated DF-3 injury at C5-C6, stabilized with four different fixation techniques: anterior plate alone, anterior plate combined with posterior wire, transarticular facet screws, and a pedicle screw-rod construct, respectively. Results. The anterior plate alone did not improve stability compared to the intact spine condition, thus allowing considerable range of motion around all three cardinal axes (p > 0.05). The anterior plate combined with posterior wire technique improved flexion-extension stiffness (p = 0.023), but not in axial rotation and lateral bending. When the anterior plate was combined with transarticular facet screws or with a pedicle screws-rod instrumentation, the stability improved in flexion-extension, lateral bending, and in axial rotation (p < 0.05). Conclusions. These findings imply that the use of anterior fixation alone is insufficient for fixation of the highly unstable DF-3 injury. In these situations, the use of anterior fixation combined with a competent posterior tension band reconstruction (e.g. transarticular screws or a posterior pedicle screws-rod device) improves segmental stability.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Ortopedi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Orthopaedics (hsv//eng)

Nyckelord

Biomechanical analysis
cervical spine
distractive-flexion injury
internal fixation
transarticular screws
pedicle screws

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Henriques, Thoma ...
Cunningham, Brya ...
Mcafee, Paul C.
Olerud, Claes
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
och Ortopedi
Artiklar i publikationen
Upsala Journal o ...
Av lärosätet
Uppsala universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy