Increased bone contact to a calcium incorporated oxidized c.p. titanium implant: an in vivo study in rabbit.
-
- Fröjd, Victoria, 1986 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper,Institute of Clinical Sciences
-
- Franke Stenport, Victoria, 1970 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för odontologi,Institute of Odontology
-
- Meirelles, Luiz, 1974 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för odontologi,Institute of Odontology
-
visa fler...
-
- Wennerberg, Ann, 1955 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper,Institute of Clinical Sciences
-
visa färre...
-
(creator_code:org_t)
- Elsevier BV, 2008
- 2008
- Engelska.
-
Ingår i: International Journal of Oral and Maxillofacial Surgery. - : Elsevier BV. - 0901-5027. ; 37:6, s. 561-566
- Relaterad länk:
-
https://gup.ub.gu.se...
-
visa fler...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- The aim of this study was to evaluate the bone response to an oxidized titanium implant (Ox) and a calcium-incorporated oxidized titanium implant (Ca). A blasted titanium implant (Bl) was used as control. The implants were topographically characterized using an optical interferometer and placed: one in each distal femoral metaphysis and two in each proximal tibial metaphysis in rabbits. The rabbits were killed 12 weeks after implant insertion, and the implants and their surrounding tissues were removed en bloc for histomorphometrical evaluations. Topographical evaluation revealed three different surfaces: average height deviation (S(a), microm) values for Ca:Ox:Bl implants were 0.3:0.6:0.9, developed surface area ratios (%) 17:44:31, number of summits per microm(2) 208:136:118, and core fluid retention index values 1.33:1.33:1.38. The mean percentages of bone contact to the implants placed in the tibia (Ca:Ox:Bl) were 47:30:34 and to the implants placed in the femur (Ca:Ox) 32:20. The mean percentages of surrounding bone area for the implants placed in the tibia were 40:47:37 and for the implants placed in the femur 43:46. A significant increase in bone contact was found for smooth (S(a) <0.5 microm) but more densely peaked calcium-incorporated oxidized implants when compared to slightly rougher (S(a)=0.5-1.0 microm) oxidized or blasted implants.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Odontologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Dentistry (hsv//eng)
Nyckelord
- titanium implants
- anodic oxidation
- calcium
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas