Intra-observer repeatability when assessing the foetal urinary bladder volume by the Virtual Organ Computer-aided AnaLysis and SUM-OF-CYLINDERS methods: A pilot study

Fagerquist, M. (författare)
NU-sjukvården,NU Hospital Group,Innlandet Hospital Trust
Sethi, Vivek (författare)
Akershus University Hospital
Skytteren, E. (författare)
Innlandet Hospital Trust
visa fler...
Odén, Anders, 1942 (författare)
Chalmers tekniska högskola,Chalmers University of Technology
visa färre...
 (creator_code:org_t)
2017
2017
Engelska.
Ingår i: Ultrasound. - 1742-271X. ; 25:3, s. 156-165
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Introduction : The aim of this study was to compare the intra-observer repeatability when using two different methods for estimating the volume of foetal urinary bladders. Method : The urinary bladders of 20 foetuses were documented by three-dimensional ultrasound. Standard deviation was compared when the volumes of identical bladder images were repeatedly estimated using the Virtual Organ Computer-aided AnaLysis and the experimental SUM-OF-CYLINDERS methods. Results : No systematic deviation was found between the estimated volumes when using these two methods. Standard deviation was smaller for the SUM-OF-CYLINDERS compared to the Virtual Organ Computer-aided AnaLysis method (p < 0.0001). In relation to bladder volumes of 5–25 ml, standard deviation was 11–14% for the Virtual Organ Computer-aided AnaLysis and 4–5% for the SUM-OF-CYLINDERS method. Conclusions : Using three-dimensional ultrasound images adapted for the Virtual Organ Computer-aided AnaLysis method, foetal urinary bladder volumes can also be estimated using the SUM-OF-CYLINDERS method. The SUM-OF-CYLINDERS method employs technical advances which may result in a lower standard deviation and therefore higher intra-observer repeatability.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
NATURVETENSKAP  -- Data- och informationsvetenskap -- Datorseende och robotik (hsv//swe)
NATURAL SCIENCES  -- Computer and Information Sciences -- Computer Vision and Robotics (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Image Processing (hsv//eng)

Nyckelord

ultrasonography
diagnostic technique and procedure
Foetus
urinary bladder
reliability

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy