Improving standard volar plate fixation in 3D-guided corrective osteotomy of the distal radius: evaluation of a shim instrument

Gryska, Emilia, 1992 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
Libberecht, Katleen, 1974 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
Swinkels, Charlotte Stor, 1987 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
visa fler...
Axelsson, Peter, 1963 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
Fredrikson, Per (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper,Institute of Clinical Sciences
Björkman, Anders (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
visa färre...
 (creator_code:org_t)
2024
2024
Engelska.
Ingår i: JOURNAL OF PLASTIC SURGERY AND HAND SURGERY. - 2000-656X .- 2000-6764. ; 59, s. 46-52
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Standard volar plates often do not fit the surface of the malunited distal radius after osteotomy, necessitating an offset angle for accurate volar tilt correction. The correction can be achieved if the plate is held at the correct angle when the distal screws are locked. With the advantage of 3D surgical planning and patient-specific instruments, we developed a shim instrument to assist the surgeon in securing the plate at the intended angle when locking the distal screws, and evaluated radiological results. Five female patients aged 63-74 with dorsally angulated extra-articular malunions underwent surgery using 3D-printed guides and the shim instrument. The plate position, drilling guide alignment, screw placements, and distal radius correction on postoperative CTs were compared with the surgical plans. Errors were measured using an anatomical coordinate system, and standard 2D radiographic measures were extracted. Preoperative dorsal tilt ranged from 16 degrees to 35 degrees, and postoperative volar tilt from 1 degrees to 11 degrees. 3D analysis revealed mean absolute correction errors of 6.1 degrees in volar tilt, 1.6 degrees in radial inclination, and 0.6 mm in ulnar variance. The volar tilt error due to the shim instrument, indicated by the mean angle error of the distal screws to the plate, was 2.1 degrees but varied across the five patients. Settling of the distal radius, due to tension during and after reduction, further contributed to a mean loss of 3.5 degrees in volar tilt. The shim instrument helped with securing plates at the intended angle; however, further correction improvements should consider the tension between the fragments of osteoporotic bone.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Ortopedi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Orthopaedics (hsv//eng)

Nyckelord

Virtual surgical planning
3D printing
patient- specific instruments
distal radius
extra-articular
osteotomy

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy