Sökning: L773:2000 656X OR L773:2000 6764
> (2020-2024) >
Improving standard ...
Improving standard volar plate fixation in 3D-guided corrective osteotomy of the distal radius: evaluation of a shim instrument
-
- Gryska, Emilia, 1992 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
-
- Libberecht, Katleen, 1974 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
-
- Swinkels, Charlotte Stor, 1987 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
-
visa fler...
-
- Axelsson, Peter, 1963 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
-
- Fredrikson, Per (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper,Institute of Clinical Sciences
-
- Björkman, Anders (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
-
visa färre...
-
(creator_code:org_t)
- 2024
- 2024
- Engelska.
-
Ingår i: JOURNAL OF PLASTIC SURGERY AND HAND SURGERY. - 2000-656X .- 2000-6764. ; 59, s. 46-52
- Relaterad länk:
-
https://gup.ub.gu.se...
-
visa fler...
-
https://doi.org/10.2...
-
visa färre...
Abstract
Ämnesord
Stäng
- Standard volar plates often do not fit the surface of the malunited distal radius after osteotomy, necessitating an offset angle for accurate volar tilt correction. The correction can be achieved if the plate is held at the correct angle when the distal screws are locked. With the advantage of 3D surgical planning and patient-specific instruments, we developed a shim instrument to assist the surgeon in securing the plate at the intended angle when locking the distal screws, and evaluated radiological results. Five female patients aged 63-74 with dorsally angulated extra-articular malunions underwent surgery using 3D-printed guides and the shim instrument. The plate position, drilling guide alignment, screw placements, and distal radius correction on postoperative CTs were compared with the surgical plans. Errors were measured using an anatomical coordinate system, and standard 2D radiographic measures were extracted. Preoperative dorsal tilt ranged from 16 degrees to 35 degrees, and postoperative volar tilt from 1 degrees to 11 degrees. 3D analysis revealed mean absolute correction errors of 6.1 degrees in volar tilt, 1.6 degrees in radial inclination, and 0.6 mm in ulnar variance. The volar tilt error due to the shim instrument, indicated by the mean angle error of the distal screws to the plate, was 2.1 degrees but varied across the five patients. Settling of the distal radius, due to tension during and after reduction, further contributed to a mean loss of 3.5 degrees in volar tilt. The shim instrument helped with securing plates at the intended angle; however, further correction improvements should consider the tension between the fragments of osteoporotic bone.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Ortopedi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Orthopaedics (hsv//eng)
Nyckelord
- Virtual surgical planning
- 3D printing
- patient- specific instruments
- distal radius
- extra-articular
- osteotomy
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas