Can scatter correction software replace a grid in DR pelvic examinations?

Precht, Helle (författare)
University College Lillebaelt
Mørup, Svea Deppe (författare)
University College Lillebaelt
Tingberg, Anders (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
visa fler...
Outzen, Claus Bjørn (författare)
University College Lillebaelt
Kusk, Kirsten Weber (författare)
Hospital of South West Jutland, Esbjerg
Nielsen, Rita Møller (författare)
Hospital of South West Jutland, Esbjerg
Midtgård, Marie (författare)
Lillebaelt Hospital
Winther, Martin Brændegaard (författare)
Slagelse Hospital
Waaler, Dag (författare)
Norwegian University of Science and Technology
Kusk, Martin Weber (författare)
Hospital of South West Jutland, Esbjerg
visa färre...
 (creator_code:org_t)
2019-05-21
2019
Engelska 9 s.
Ingår i: Radiation Protection Dosimetry. - : Oxford University Press (OUP). - 1742-3406 .- 0144-8420. ; 187:1, s. 8-16
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The purpose was to examine if scatter correction software could replace a grid while maintaining image quality and reducing radiation dose for pelvic DR examinations. Grid images was produced with 70 kV and 16mAs. Anthropomorphic- and Contrast Detail RADiography (CDRAD) non-grid images were produced with 60 kV, 80 kV and 90 kV combined with five different mAs and scatter correction software. The anthropomorphic images were analyzed by absolute Visual Grading Analysis (VGA). The CDRAD images were analyzed using the CDRAD analysis software. The results showed a total of 54.6% non-grid images were evaluated as unsuitable for diagnostic use by the VGA. The CDRAD grid images showed that the IQF_inv values were significantly different (p = 0.0001) when compared to every group of non-grid images. Hereby, the conclusion stated that the scatter correction software did not compensate for the loss in image quality due to scattered radiation at the exposure levels included in a pelvic examination.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Digital radiography
image quality
pelvic
Radiation dose
Scatter correction software

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy