100a Bäck, S Au Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups,Skåne University Hospital4 aut0 (Swepub:lu)rfa-sba
2451 0a Verification of single beam treatment planning using a ferrous dosimeter gel and MRI (FeMRI)
264 1c 1998
300 a 6 s.
520 a A method for analysing and comparing treatment planning system (TPS) data and ferrous dosimeter gel measurements evaluated with MRI (FeMRI) was developed, including image processing to final absorbed dose images. Measurements were analysed according to this method and FeMRI data were thereby compared with the TPS-calculated dose distribution. For photons, differences between FeMRI- and TPS dose data were mainly within +/- 2%. Minor shortcomings found in both the FeMRI system and the TPS are explained and discussed. For electron beams, there was an overall good agreement. It was found that the TPS underestimates the lateral scattering dose outside the primary beam, but the reported dose difference corresponds to a small spatial deviation (less than 2 mm). It is important to consider this single beam data comparison when the method is extended to more complicated situations, for example when using several beams.
650 7a MEDICIN OCH HÄLSOVETENSKAPx Klinisk medicinx Radiologi och bildbehandling0 (SwePub)302082 hsv//swe
650 7a MEDICAL AND HEALTH SCIENCESx Clinical Medicinex Radiology, Nuclear Medicine and Medical Imaging0 (SwePub)302082 hsv//eng
653 a Ferrous Compounds
653 a Gels
653 a Humans
653 a Magnetic Resonance Imaging
653 a Radiotherapy Dosage
653 a Reproducibility of Results
653 a Sensitivity and Specificity
700a Magnusson, Peteru Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups4 aut0 (Swepub:lu)extLU-696
700a Olsson, L Eu Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups4 aut0 (Swepub:lu)rfa-lol
700a Montelius, A4 aut
700a Fransson, A4 aut
700a Mattsson, Su Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups4 aut0 (Swepub:lu)rfa-sma
710a Medicinsk strålningsfysik, Malmöb Forskargrupper vid Lunds universitet4 org
773t Acta Oncologicag 37:6, s. 6-561q 37:6<6-561x 0284-186X
8564 8u https://lup.lub.lu.se/record/866cd78a-5311-418d-ae62-4b09781b611b

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Bäck, S A
Magnusson, Peter
Olsson, L E
Montelius, A
Fransson, A
Mattsson, S
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
och Radiologi och bi ...
Artiklar i publikationen
Acta Oncologica
Av lärosätet
Lunds universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy