A new method of intrathecal PO2, PCO2, and pH measurements for continuous monitoring of spinal cord ischemia during thoracic aortic clamping in pigs

Christiansson, Lennart (författare)
Uppsala universitet,Institutionen för kirurgiska vetenskaper,Anaesthesiology and Intensive Care
Hellberg, Anders (författare)
Uppsala universitet,Institutionen för kirurgiska vetenskaper,Vascular Surgery
Koga, Itaru (författare)
Uppsala universitet,Institutionen för kirurgiska vetenskaper,Anaesthesiology and Intensive Care
visa fler...
Thelin, Stefan (författare)
Uppsala universitet,Institutionen för kirurgiska vetenskaper,Thoracic Surgery
Bergqvist, David (författare)
Uppsala universitet,Kärlkirurgi
Wiklund, Lars (författare)
Uppsala universitet,Anestesiologi och intensivvård
Karacagil, Sadettin (författare)
Uppsala universitet,Institutionen för kirurgiska vetenskaper,Vascular Surgery
visa färre...
 (creator_code:org_t)
Elsevier BV, 2000
2000
Engelska.
Ingår i: Surgery. - : Elsevier BV. - 0039-6060 .- 1532-7361. ; 127:5, s. 571-576
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • BACKGROUND: Impaired spinal cord circulation during thoracic aortic clamping may result in paraplegia. Reliable and fast responding methods for intraoperative monitoring are needed to facilitate the evaluation of protective measures and efficiency of revascularization.METHODS: In 11 pigs, a multiparameter PO2, PCO2, and pH sensor (Paratrend 7, Biomedical Sensors Ltd, United Kingdom) was introduced into the intrathecal space for continuous monitoring of cerebrospinal fluid (CSF) oxygenation during thoracic aortic cross-clamping (AXC) distal to the left subclavian artery. A laser-Doppler probe was inserted into the epidural space for simultaneous measurements of spinal cord flux. Registrations were made before and 30 minutes after clamping and 30 and 60 minutes after declamping. The same measuring points were used for systemic hemodynamic and metabolic data acquisition.RESULTS: The mean CSF PO2 readings of 41 mm Hg (5.5 kPa) at baseline decreased within 3 minutes to 5 mm Hg (0.7 kPa) during AXC (P < .01). Spinal cord flux measurement responded immediately in the same way to AXC. Both methods indicated normalization of circulation during declamping. Significant (P < .01) changes were also observed in the CSF metabolic parameters PCO2 and pH.CONCLUSIONS: In this experimental model of spinal ischemia by AXC, online monitoring of intrathecal PO2, PCO2, and pH showed significant changes and correlated well with epidural laser-Doppler flowmetry (P < .01).

Nyckelord

MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

  • Surgery (Sök värdpublikationen i LIBRIS)

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy