Technical evaluation of different motion-monitoring systems for respiratory gating in radiation therapy

Jönsson, Mattias (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Nordström, Fredrik (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Kugele, Malin (author)
show more...
Ljungqvist, Rebecca (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Thornberg, Charlotte (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Bäck, Sven (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Ceberg, Sofie (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
show less...
 (creator_code:org_t)
2013
2013
English.
In: Medical Physics in the Baltic States. - 1822-5721. ; , s. 38-41
  • Conference paper (peer-reviewed)
Abstract Subject headings
Close  
  • The purpose of this work was to evaluate three motion-monitoring systems: Sentinel (TM), Catalyst (TM) (both C-RAD) and Real-time Position Management (RPM (TM)) system (Varian Medical Systems Inc) for respiratory gating in radiotherapy. To measure the systems. latency, an in-house built circuit with a microcontroller operating a piston was used. The trigger pulse was sent from the gating systems to the accelerator within 300 mu s. However, the response time of the accelerator could be over 300 ms

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Keyword

Breathing adapted radiotherapy
respiratory gating
optical surface
scanning

Publication and Content Type

kon (subject category)
ref (subject category)

Find in a library

To the university's database

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view