SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Mohammed Abbas)
 

Sökning: WFRF:(Mohammed Abbas) > Citrate-selective e...

Citrate-selective electrochemical mu-sensor for early stage detection of prostate cancer

Azzouzi, Sawsen (författare)
University of Sousse, Tunisia
Patra, Hirak Kumar (författare)
Linköpings universitet,Avdelningen för cellbiologi,Medicinska fakulteten
Ben Ali, Mounir (författare)
University of Sousse, Tunisia
visa fler...
Nooredeen Abbas, Mohammed (författare)
National Research Centre, Egypt
Dridi, Cherif (författare)
Centre Research Microelect and Nanotechnol CRMN Sousse, Tunisia
Errachid, Abdelhamid (författare)
University of Lyon 1, France
Turner, Anthony (författare)
Linköpings universitet,Biosensorer och bioelektronik,Tekniska fakulteten
visa färre...
 (creator_code:org_t)
ELSEVIER SCIENCE SA, 2016
2016
Engelska.
Ingår i: Sensors and actuators. B, Chemical. - : ELSEVIER SCIENCE SA. - 0925-4005 .- 1873-3077. ; 228, s. 335-346
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The extremely specialised anatomical function of citrate inside the prostate, make it one of the preferred biomarkers for early stage detection of prostate cancer. However, current detection methods are seriously limited due to the very low citrate concentrations that need to be measured in order to follow disease progression. In the present work, we report a novel citrate-selective-sensor based on iron (III) phthalocyanine chloride-C-monoamido-Poly-n-Butyl Acrylate (Fe(III)MAPcC1 P n BA) modified gold -electrodes for the electrochemical determination and estimation of the pathophysiological range of citrate. The newly synthesised ionophore has been structurally characterised using Fourier transform infrared (FTIR) and UV-vis spectroscopy. Contact angle measurements and atomic force microscopy (AFM) have been used to investigate the adhesion and morphological properties of the membrane. The developed citrate-selective-electrodes had a Nernstian sensitivity of-19.34 +/- 0.83 mV/decade with a detection limit of about 9 x 10-6M and a linear range from 4 x 10(-5)M to 10(-1) M, which covered the pathologically important clinical range. Electrochemical impedance spectroscopy (EIS) showed very high sensitivity with a lower Limit of detection 1.7 x 10(-9) M and linear detection range (10(-8)-10(-1) M), which is very important not only for the early-stage diagnosis and screening procedures, but also in mapping the stage of the cancer too. (C) 2016 Elsevier B.V. All rights reserved.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk laboratorie- och mätteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Laboratory and Measurements Technologies (hsv//eng)
NATURVETENSKAP  -- Kemi -- Analytisk kemi (hsv//swe)
NATURAL SCIENCES  -- Chemical Sciences -- Analytical Chemistry (hsv//eng)

Nyckelord

Citrate selective ionophore; Potentiometry; Impedance spectroscopy; Prostate cancer; Early-stage detection

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy