CHICSi - a compact ultra-high vacuum compatible detector system for nuclear reaction experiments at storage rings. II. Detectors
-
- Golubev, Pavel (författare)
- Lund University,Lunds universitet,Kärnfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Nuclear physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
-
- Avdeichikov, Vladimir (författare)
- Lund University,Lunds universitet,Kärnfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Nuclear physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
-
- Carlén, Lars (författare)
- Lund University,Lunds universitet,Kärnfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Nuclear physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
-
visa fler...
-
- Jakobsson, Bo (författare)
- Lund University,Lunds universitet,Kärnfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Nuclear physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
-
Siwek, A (författare)
-
van Veldhuizen, EJ (författare)
-
Westerberg, L (författare)
-
- Whitlow, Harry J (författare)
- Lund University,Lunds universitet,Kärnfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Nuclear physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
-
visa färre...
-
(creator_code:org_t)
- 2003
- 2003
- Engelska.
-
Ingår i: Nuclear Instruments & Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment. - 0167-5087. ; 500:1-3, s. 96-115
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- We describe the detectors for identification of charged particles and fragments in CHICSi, a large solid angle multitelescope system mounted inside an ultra-high vacuum (UHV), cluster-jet target chamber. CHICSi performs nuclear reaction experiments at storage rings. The telescopes consist of a first very thin, 10-14 mum Si detector, a second 300 mum (or possibly 500 pm) ion implanted Si detector supplemented by a 6 mm GSO(Ce) scintillator read out by a photodiode (PD) or by a third 300 mum Si detector. The telescopes provide full charge separation up to Z = 17 and mass resolution up to A = 9 in the energy range 0.7-60A MeV. The thin p-i-n diode detector, etched out from a 280 mum Si wafer, and the GSO/PD detector, both exclusively developed for CHICSi, provide an energy resolution less than or equal to 8%, while the standard 300 mum detectors have less than or equal to 2% energy resolution. Radiation stability of the Si detectors is confirmed up to an integrated flux of 10(10) alpha particles. The GSO detector has 70% light collection efficiency with the optical coupling to the PD a simple open, 0.2 mm, gap. A new method, developed to perform absolute energy calibration for the GSO/PD detector is presented. (C) 2003 Elsevier Science B.V. All rights reserved.
Ämnesord
- NATURVETENSKAP -- Fysik -- Acceleratorfysik och instrumentering (hsv//swe)
- NATURAL SCIENCES -- Physical Sciences -- Accelerator Physics and Instrumentation (hsv//eng)
Nyckelord
- multi-detector system
- ultra-high vacuum compatibility
- ultra-thin
- detectors