Breast tomosynthesis: Accuracy of tumor measurement compared with digital mammography and ultrasonography.

Förnvik, Daniel (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Zackrisson, Sophia (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
Ljungberg, Otto (författare)
Lund University,Lunds universitet,Patologi, Malmö,Forskargrupper vid Lunds universitet,Pathology, Malmö,Lund University Research Groups
visa fler...
Svahn, Tony (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Timberg, Pontus (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Medicinsk strålningsfysik, Malmö,Radiology Diagnostics, Malmö,Lund University Research Groups,Medical Radiation Physics, Malmö
Tingberg, Anders (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Andersson, Ingvar (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
visa färre...
 (creator_code:org_t)
2010-04-01
2010
Engelska.
Ingår i: Acta radiologica (Stockholm, Sweden : 1987). - : SAGE Publications. - 1600-0455 .- 0284-1851. ; 51, s. 240-247
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: Mammographic tumor size measurement can be difficult because breast structures are superimposed onto a two-dimensional (2D) plane, potentially obscuring the tumor outline. Breast tomosynthesis (BT) is a 3D X-ray imaging technique in which low-dose images are acquired over a limited angular range at a total dose comparable to digital mammography (DM). These low-dose images are used to mathematically reconstruct a 3D image volume of the breast, thus reducing the problem of superimposed tissue. Purpose: To investigate whether breast cancer size can be more accurately assessed with breast tomosynthesis than with digital mammography and ultrasonography (US), by reducing the disturbance effect of the projected anatomy. Material and Methods: A prototype BT system was used. The main inclusion criterion for BT examination was subtle but suspicious findings of breast cancer on 2D mammography. Sixty-two women with 73 breast cancers were included. BT, DM, and US sizes were measured independently by experienced radiologists without knowledge of the pathology results, which were used as reference. Results: The tumor outline could be determined in significantly more cases with BT (63) and US (60) than DM (49). BT and US size correlated well with pathology (R=0.86 and R=0.85, respectively), and significantly better than DM size (R=0.71). Accordingly, staging was significantly more accurate with BT than with DM. Conclusion: The study indicates that BT is superior to DM in the assessment of breast tumor size and stage.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy