MR-IMPACT : comparison of perfusion-cardiac magnetic resonance with single-photon emission computed tomography for the detection of coronary artery disease in a multicentre, multivendor, randomized trial

Schwitter, Juerg (författare)
Wacker, Christian M. (författare)
van Rossum, Albert C. (författare)
visa fler...
Lombardi, Massimo (författare)
Al-Saadi, Nidal (författare)
Ahlström, Håkan, 1953- (författare)
Uppsala universitet,Enheten för radiologi,Ahlström, H.
Dill, Thorsten (författare)
Larsson, Henrik B. W. (författare)
Flamm, Scott D. (författare)
Marquardt, Moritz (författare)
Johansson, Lars (författare)
Uppsala universitet,Enheten för radiologi,Ahlström, H
visa färre...
 (creator_code:org_t)
2008-01-21
2008
Engelska.
Ingår i: European Heart Journal. - : Oxford University Press (OUP). - 0195-668X .- 1522-9645. ; 29:4, s. 480-9
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • AIMS: To determine in a multicentre, multivendor trial the diagnostic performance for perfusion-cardiac magnetic resonance (perfusion-CMR) in comparison with coronary X-ray angiography (CXA) and single-photon emission computed tomography (SPECT). METHODS AND RESULTS: Of 241 eligible patients from 18 centres, 234 were randomly dosed with 0.01, 0.025, 0.05, 0.075, or 0.1 mmol/kg Gd-DTPA-BMA (Omniscantrade mark, GE-Healthcare) per stress (0.42 mg/kg adenosine) and rest perfusion study. Coronary artery disease (CAD) was defined as diameter stenosis > or =50% on quantitative CXA. Five CMR and eight SPECT studies (of 225 complete studies) were excluded from analyses due to inadequate quality (three blinded readers scored per modality). The comparison of CMR vs. SPECT was based on receiver operating characteristic (ROC) analysis. Perfusion-CMR at the optimal CM dose (0.1 mmol/kg) had similar performance as SPECT, if only the SPECT studies of the 42 patients with this dose were considered [area under ROC curve (AUC): 0.86 +/- 0.06 vs. 0.75 +/- 0.09 for SPECT, P = 0.12]; however, diagnostic performance of perfusion-CMR was better vs. the entire SPECT population (AUC: 0.67 +/- 0.05, n = 212, P = 0.013). CONCLUSIONS: In this multicentre, multivendor trial, ROC analyses suggest perfusion-CMR as a valuable alternative to SPECT for CAD detection showing equal performance in the head-to-head comparison. Comparing perfusion-CMR with the entire SPECT population suggests CMR superiority over SPECT, which warrants further evaluation in larger trials.

Nyckelord

MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy