Determining Heart Activity Present in the Pressure Sensors of a Dialysis Machine

Holmer, Mattias (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Grigonyte, Egle (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Solem, Kristian (författare)
Lund University,Lunds universitet,Institutioner vid LTH,Lunds Tekniska Högskola,Departments at LTH,Faculty of Engineering, LTH
visa fler...
Olde, Bo (författare)
Sandberg, Frida (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Sörnmo, Leif (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Murray, Alan (redaktör/utgivare)
visa färre...
 (creator_code:org_t)
2013
2013
Engelska.
Ingår i: Computing in Cardiology. - 2325-8861. ; , s. 217-220
  • Konferensbidrag (refereegranskat)
Abstract Ämnesord
Stäng  
  • Determination of heart status during dialysis can im- prove patient monitoring. Pressure sensors in the dialysis machine measures the heart pulses that propagates in the body and enter the extracorporeal blood circuit. A peri- staltic blood pump, located in the same circuit, introduces strong periodic pressure pulses that interfere with the much weaker cardiac component. These signal characteristics make the extraction of the heart activity challenging. In the present study, we explore the possibility to extract and analyze the cardiac component using simulated data. The accuracy of the timing of each heartbeat is analyzed. Ad- ditionally, the heart component is extracted from patient pressure recordings, and compared to the heart rate com- puted from a photoplethysmogram. The results show that heart timings can be accurately determined using the pres- sure sensors of a dialysis machine.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering (hsv//eng)

Publikations- och innehållstyp

kon (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy