Evaluation of two commercial CT metal artifact reduction algorithms for use in proton radiotherapy treatment planning in the head and neck area

Andersson, Karin M., 1989- (författare)
Örebro universitet,Institutionen för medicinska vetenskaper,The Skandion Clinic, Uppsala, Sweden,Skand Clin, Uppsala, Sweden; Örebro Univ, Sch Hlth & Med Sci, Örebro, Sweden
Vallhagen Dahlgren, Christina (författare)
The Skandion Clinic, Uppsala, Sweden,Skand Clin, Uppsala, Sweden
Reizenstein, Johan (författare)
Department of Oncology, Faculty of Medicine and Health, Örebro University, Örebro, Sweden,Örebro Univ, Fac Med & Hlth, Dept Oncol, Örebro, Sweden
visa fler...
Cao, Yang, Associate Professor, 1972- (författare)
Örebro universitet,Institutionen för medicinska vetenskaper,Region Örebro län,Unit of Biostatistics, Institute of Environmental Medicine, Karolinska Institutet, Stockholm, Sweden,Clinical Epidemiology and Biostatistics,Örebro Univ, Sch Med Sci, Clin Epidemiol & Biostat, Örebro, Sweden; Karolinska Inst, Inst Environm Med, Unit Biostat, Stockholm, Sweden
Ahnesjö, Anders, 1953- (författare)
Uppsala universitet,Medicinsk strålningsvetenskap
Thunberg, Per, 1968- (författare)
Örebro universitet,Institutionen för medicinska vetenskaper,Department of Medical Physics,Örebro Univ, Fac Med & Hlth, Dept Med Phys, Örebro, Sweden
visa färre...
 (creator_code:org_t)
2018-09-19
2018
Engelska.
Ingår i: Medical physics (Lancaster). - : Wiley-Blackwell Publishing Inc.. - 0094-2405 .- 2473-4209. ; 45:10, s. 4329-4344
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • PURPOSE: To evaluate two commercial CT metal artifact reduction (MAR) algorithms for use in proton treatment planning in the head and neck (H&N) area.METHODS: An anthropomorphic head phantom with removable metallic implants (dental fillings or neck implant) was CT-scanned to evaluate the O-MAR (Philips) and the iMAR (Siemens) algorithms. Reference images were acquired without any metallic implants in place. Water equivalent thickness (WET) was calculated for different path directions and compared between image sets. Images were also evaluated for use in proton treatment planning for parotid, tonsil, tongue base, and neck node targets. The beams were arranged so as to not traverse any metal prior to the target, enabling evaluation of the impact on dose calculation accuracy from artifacts surrounding the metal volume. Plans were compared based on γ analysis (1 mm distance-to-agreement/1% difference in local dose) and dose volume histogram metrics for targets and organs at risk (OARs). Visual grading evaluation of 30 dental implant patient MAR images was performed by three radiation oncologists.RESULTS: In the dental fillings images, ΔWET along a low-density streak was reduced from -17.0 to -4.3 mm with O-MAR and from -16.1 mm to -2.3 mm with iMAR, while for other directions the deviations were increased or approximately unchanged when the MAR algorithms were used. For the neck implant images, ΔWET was generally reduced with MAR but residual deviations remained (of up to -2.3 mm with O-MAR and of up to -1.5 mm with iMAR). The γ analysis comparing proton dose distributions for uncorrected/MAR plans and corresponding reference plans showed passing rates >98% of the voxels for all phantom plans. However, substantial dose differences were seen in areas of most severe artifacts (γ passing rates of down to 89% for some cases). MAR reduced the deviations in some cases, but not for all plans. For a single patient case dosimetrically evaluated, minor dose differences were seen between the uncorrected and MAR plans (γ passing rate approximately 97%). The visual grading of patient images showed that MAR significantly improved image quality (P < 0.001).CONCLUSIONS: O-MAR and iMAR significantly improved image quality in terms of anatomical visualization for target and OAR delineation in dental implant patient images. WET calculations along several directions, all outside the metallic regions, showed that both uncorrected and MAR images contained metal artifacts which could potentially lead to unacceptable errors in proton treatment planning. ΔWET was reduced by MAR in some areas, while increased or unchanged deviations were seen for other path directions. The proton treatment plans created for the phantom images showed overall acceptable dose distributions differences when compared to the reference cases, both for the uncorrected and MAR images. However, substantial dose distribution differences in the areas of most severe artifacts were seen for some plans, which were reduced by MAR in some cases but not all. In conclusion, MAR could be beneficial to use for proton treatment planning; however, case-by-case evaluations of the metal artifact-degraded images are always recommended.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

computed tomography
dose calculation
metal artifacts
proton therapy
radiotherapy

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy